Validation of Electrical Test
dhaas is testing TDR2 number 100, FM
-> AMSWire/CDP Test
Power consumptions
Volt | Power (mA) | Volt | Power [mA] |
A +3.3 | 110.00000 | B +3.3 | 109.00000 |
A +5.6S | 88.00000 | B +5.6S | 96.00000 |
A +5.6K | 53.00000 | B +5.6K | 56.00000 |
A -2.0S | 10.00000 | B -2.0S | 9.00000 |
A -2.0K | 5.00000 | B -2.0K | 5.00000 |
A +2.0S | 0.00000 | B +2.0S | 0.00000 |
A +2.0K | 0.00000 | B +2.0K | 0.00000 |
Onboard Voltages
Volt nom. | Onboard Volt | Volt nom. | Onboard Volt |
A +3.3 | 3.413 | B +3.3 | 3.428 |
A +5.6S | 5.193 | B +5.6S | 5.216 |
A +5.6K | 5.214 | B +5.6K | 5.220 |
A -2.0K | -2.011 | B -2.0K | -2.015 |
A +2.0K | 2.159 | B +2.0K | 2.155 |
A -2.0S | 3.411 | B -2.0S | -2.013 |
A +2.0S | 3.428 | B +2.0S | 2.159 |
SSF Voltages
SSF | Cut Value (mA) | SSF | Cut Value (mA) |
A +3.3 | 439.0 | B +3.3 | 401.0 |
A +5.6S | 0 | B +5.6S | 0 |
A +5.6K | 114.0 | B +5.6K | 122.0 |
Comments:
Voltage +2.0 S, Side A is ok (+2.2 V)
Voltage -2.0 S, Side A is ok (-2.0 V) (went to the wrong testpoints first)
SSF +5.6 S, Side B is ok, cuts at 141 mA
SSF +5.6 S, Side A is ok, cuts at 142 mA
Data files of this test:
./FM_100_20070119_test1.log
./FM_100_20070119_test1_ssf.eps
./FM_100_20070119_test1_ssf1.dat
./FM_100_20070119_test1_ssf2.dat
./FM_100_20070119_test1_ssf3.dat
./FM_100_20070119_test1_ssf4.dat
./FM_100_20070119_test1_ssf5.dat
./FM_100_20070119_test1_ssf6.dat